Study evaluated the accuracy, reliability, and discriminatory ability of the HipScreen mobile app for measuring hip migration percentage in children with cerebral palsy. 37 individuals, including 30 healthcare professionals, used the app to measure 40 hips, showing a mean absolute error of 5.72% and sensitivity of 94%. Researchers suggest this technology can be used for effective hip surveillance across diverse user backgrounds.